GEHMAN’S TORSION TESTER

GEHMAN’S TORSION TESTER

(1 đánh giá)

TM-2535

Ueshima Seisakusho

12 months

Rubber loses resiliency as temperature goes down. The frozen rubber can't keep its essential characteristics under the temperature below Glass Transition Point. In fact, the rubber applications under severe environments such as vehicle tire in frozen area and aircraft components in the stratosphere are seriously affected. Our Gehman's torsion tester TM-2535 is the most popular testing machine for this purpose and is adopted by JIS K6261(Equivalent to ISO 1432, ASTM D1053).

Rubber  loses  resiliency as  temperature   goes  down. The frozen rubber can't keep  its essential characteristics under the temperature  below Glass Transition Point.
In fact,  the rubber applications under severe environments such  as vehicle tire in frozen  area  and aircraft components  in the  stratosphere  are  seriously affected.
It is a vital importance with the rubber applicants to test low temperature  characteristics of vulcanized rubber.
Our Gehman's torsion tester is the most popular testing machine for this purpose  and  is  adopted by  JIS K6261(Equivalent to ISO 1432) for low temperature testing method of vulcanized  rubber.
The Model   TM-2530  satisfies performance requirements specified by the standard JIS K6261,  and equips with the exclusive use frozen machine.
It is also designed to perform measurement and data processing in a full automatic for easier operation.
○ Simultaneous  tests of  6 specimens  including one   specimen which  twist/low temperature characteristics  is known.    The testing capability of this  is recommended by the standard  JIS K6261 for reliable test data.
○ Temperature-apparent torsion nal  modulus  of  rigidity  curve  is  also  displayed  and  recorded in addition to temperature-twist angle  curve.
○The computer calculates not only relative modulus but also glass transition temperature(Tg)  which is an important measure of low temperature  characteristics of rubber.
Model TM-2535
Applicable Standard: JIS K 6261, ISO 1432
Cooling  Method: Refrigerator
Number of Samples: 6 (simultaneous measurement)
Temp. Range: -70°C   –   RT(23°C±2°C)
Temp. Rising Every 5°C (kept at each set temp. for 5 min.)
Temp. Control PID controlled digital temperature controller, sensor   PT100Ω
Heat Transfer Medium: Ethanol
Torsion Angle Setting Rotary encoder, minimum angle  display: 0.36°
Data Processing: Data (glass transition temp. Tg) display Graph (temp. – torsion angle curve, temp. – apparent torsion modulus curve) Filing (data, test conditions,  PID setting)
Release  of torsion wire
Safety Device leakage breaker, overheat protector
Power Supply 200VAC, 3 phases, 30A, 50/60Hz (100VAC for PC) or specified voltage
Dimensions (main body) (W)520x(D)780x(H)1505   mm

Đánh giá sản phẩm

Bình chọn sản phẩm:

Bình luận

Top

  Zalo