Supercritical Fluids Extraction

(10 đánh giá)

SE-600–10000

TST - Taiwan

12 months

TST Model SE-600–10000 Supercritical Fluids Extraction Machine

Supercritical Fluid Preparative Scale CO₂ Extraction Equipment
Model No.: SE-600–10000 Series

Manufacturer: TST - Taiwan

Specifications

1. System Specifications

  • Maximum Pressure: 7250 Psi / 500 Bar
  • Maximum Temperature: 100 °C

2. Main Unit Specifications

Extraction Vessel: 1 L × 1 set

  • Maximum Pressure: 7250 Psi / 500 Bar
  • Maximum Temperature: 100 °C
  • Volume: 600 ml × 1 (inner basket)
  • Material: Stainless steel
  • Heating System: Electric heating

Collection System: 1 set

  • CO₂ flow rate control valve × 1 set
  • Control valve heating module × 1 set
  • Designed for atmospheric collection

Pressure Control System: PM-10000C × 1 set
Cooling System: 1 set

3. Construction Materials

  • Piping: SUS 304
  • Frame Structure: Aluminum

4. Control System and Monitoring

  • Control Valve: Manual type / High-pressure valve
  • Temperature Control: Temperature controller with thermocouple
  • Pressure Control: Pressure gauge / Pressure transducer
  • CO₂ Flow Control: Flow meter with digital display of CO₂ gas flow rate

5. Safety and Protection

  • Rupture Disk: Set pressure 8500 Psig
  • Heating Zone: Equipped with insulation jacket

6. Utility Requirements (to be supplied by user)

  • Power Supply: 220 V, 12 A
  • Compressed Air: 7 kgf/cm²; 4 m³/min
  • CO₂ Supply: Liquid CO₂ cylinder with internal tube

7. Optional Components (non-standard)

  • Spare Parts: (e.g., seals, rupture disks, etc.)
  • Co-solvent Pump Module

Digital Pressure Control Module (PM-10000C)

Applied Fluid

  • Liquid Carbon Dioxide (LCO₂)

Specifications

  1. System Output Pressure: 1000–10000 psi (depending on driving pressure)
  2. System Output Flow Rate: 200 ml/min (up to 4350 psi)
  3. Booster Pump: Included
  4. Pre-cooler: Built-in (no external chiller required)
  5. Piping: Seamless stainless steel tubing
  6. Certification: CE compliant; EN 61010-2-010 (2nd Ed) and EN 61010-1:2001
  7. Dimensions: 360 mm × 600 mm × 270 mm
  8. Weight: 30 kg

Control System

  1. Color touch-panel for pressure control and human-machine interface (HMI)
  2. Programmable pressure control system
  3. Proportional precise pressure regulation
  4. Multi-step booster with instant adjustment / response control
  5. Alarm and timer for abnormal operation

Safety Protection

  1. Overpressure alarm system
  2. Emergency stop and automatic gas cut-off for overpressure
  3. Safety relief valve for full-flow overpressure protection

Utility Requirements

  1. Power Supply: 110 / 220 V
  2. Source Air: 5–8 kg/cm²
  3. Chiller Water Temperature: ≤ 5 °C