Product
Dịch vụ
Applied Separations
Online support
Tel: 0908866933 - secovina@vnn.vn
Thống kê truy cập
- Đang online 0
- Hôm nay 0
- Hôm qua 0
- Trong tuần 0
- Trong tháng 0
- Tổng cộng 0
TR Tester
TM-3525
Ueshima Japan
12 months
This device measures the retraction rate of elongation when a specimen is subjected to a certain elongation, frozen, and then gradually warmed up. The encoder is used to measure the retraction, and a PC is used to set test conditions, display data (TR value, retraction rate), draw graphs (temperature, retraction rate), and save the results to a file. Standard JIS K 6261-4,ISO2921,ASTM D 1329.
This device measures the retraction rate of elongation when a specimen is subjected to a certain elongation, frozen, and then gradually warmed up. The encoder is used to measure the retraction, and a PC is used to set test conditions, display data (TR value, retraction rate), draw graphs (temperature, retraction rate), and save the results to a file.
- Equipped with a dual-stage refrigeration system, ethanol can be cooled down to -73°C. The chiller can be operated on a timer, so testing can be started immediately after the time is set in advance.
- The data measured by the testing machine are linked to the PC, the test data can be automatically aggregated, and the results can be displayed.
Standard JIS K 6261-4, ISO2921, ASTM D 1329
Number of Samples: 6
Sample Size: Length of parallel part: 50 mm or 100 mm
Temp. Range: -70°C to 20°C
Test Stroke: 200 mm max. (for the 50-mm samples) *Initial elongation setting is manual.
Displacement : Rotary Encoder (2000 P/R)
Measurement: 0.065 mm / P
Cooling Method: Dual Timer -Stage (99 h Refrigerator 59 min) (3 Three-phase AC200V 1.1KW+1.1KW)
Heat Transfer Medium: Ethanol (approx. 18L)
Heating Method: Heater, 3KW thyristor control
Temperature Control: PID-controlled digital temperature controller, Sensor PT100Ω
Safety Device: Leakage breaker, overheat protector, float switch
PC Specifications:
・ OS : Windows 10
・ Slot : Three PCI slots required
・ Monitor resolution : 1280×1024 or higher
Data Processing and control
・ Equipment operation
・ Test condition setup
・ Data display (TR values, retraction rate)
・ Graphic display (temperature - retraction)
・ Filing (Data, Test Conditions, PID Settings)
Weight 210 Kg(without heat transfer medium)
Power Supply : Three-phase AC200V or 380V, 50/60Hz
Outer Dimensions: 520(W)×780(D)×1790(H)mm
Bình luận